AutoGineer summary
Written by AutoGineer from the skill’s own files, not by the author
This skill provides a structured approach to using MATLAB Audio Toolbox for acoustic measurement and analysis. It guides users through selecting the correct functions for psychoacoustic metrics, speech intelligibility, noise compliance, and broadcast loudness. The skill emphasizes proper calibration techniques, standard-compliant workflows, and the correct handling of signal data to ensure accurate results, while explicitly warning against the use of nonexistent or licensed-restricted functions.
What it does
- Calculates psychoacoustic metrics like loudness, sharpness, roughness, and fluctuation strength.
- Evaluates speech quality and intelligibility using metrics such as ViSQOL, STOI, STI, and SII.
- Assesses noise compliance using NC, RC, and RNC ratings for various environments.
- Performs offline SPL measurements and broadcast loudness checks (EBU R 128).
- Provides guidance on microphone calibration and system-level calibration factors.
- Offers standard-compliant filtering and unit conversion utilities.
When to use it
- When you need to measure perceived loudness or sound quality in a research or engineering context.
- When evaluating room acoustics or HVAC noise compliance against industry standards.
- When analyzing speech intelligibility for communication systems or public address setups.
- When you need to calibrate audio recordings for accurate SPL or metric computation.
- When you need to perform broadcast loudness compliance checks.
Advantages
- Ensures adherence to international standards like ISO 532-1/2, ANSI/ASA S12.2, and IEC 60268-16.
- Prevents common errors such as incorrect function usage, argument ordering, or manual implementation of complex algorithms.
- Provides clear guidance on calibration to ensure measurement accuracy.
- Streamlines the selection of appropriate metrics based on specific application goals.
Usage examples
- “How do I calculate the loudness of a recording in sones?”
- The agent will guide you to use the acousticLoudness function, specifying the ISO 532-1 method, applying your calibration factor, and selecting the appropriate sound field.
- “Is my office HVAC noise compliant?”
- The agent will recommend using the roomCriteria function to obtain an RC rating, explaining how to interpret the output and verify the measurement conditions.
- “How can I measure the speech quality of a processed audio file?”
- The agent will instruct you to use the visqol function with the 'speech' mode, ensuring the processed signal is passed as the first argument.
- “How do I calibrate my microphone for SPL measurements?”
- The agent will explain how to use calibrateMicrophone with a 1 kHz tone recording and a known SPL reference to compute the correct calibration factor.
Install
Download the skill as a ZIP:
https://autogineer.com/agent-skills/matlab/matlab-evaluate-acoustic-metrics/downloadUnzip it into the cross-client skills folder of your project:
mkdir -p .agents/skills && unzip matlab-evaluate-acoustic-metrics.zip -d .agents/skills
Documentation by MathWorks
Evaluate Acoustic Metrics
Select the correct Audio Toolbox function for acoustic measurement, apply proper calibration, and choose appropriate standards — instead of reimplementing algorithms or calling nonexistent functions.
When to Use
Computing psychoacoustic metrics (loudness, sharpness, roughness, fluctuation)
Evaluating speech quality or intelligibility (ViSQOL, STOI, STI, SII)
Assessing noise compliance (NC rating, RC rating, RNC, SIL)
Measuring SPL from recordings (offline, calibrated)
Computing broadcast loudness (EBU R 128 / LUFS)
Measuring reverberation time (RT60) from impulse responses
Detecting tonal content (tone-to-noise ratio, prominence ratio)
Converting microphone sensitivity to CalibrationFactor
User mentions PESQ or POLQA (redirect to correct alternatives)
When NOT to Use
Real-time streaming audio processing loops — use
matlab-process-streaming-audioAudio device I/O, recording, or live
calibrateMicrophone— usematlab-play-record-audioAudio plugin authoring (VST/AU) — use
matlab-write-audio-pluginGeneral filter design unrelated to measurement — use
matlab-design-digital-filterSpatial audio / HRTF / SOFA processing
Workflow
Follow this decision tree to route the user's goal to the right function:
1. Identify the measurement goal
| Goal | Category | Go to |
|---|---|---|
| "How loud is it?" (perceived) | Psychoacoustic | → Psychoacoustic Metrics |
| "Does it comply with broadcast loudness?" | Broadcast | → Broadcast Loudness |
| "Is the noise level acceptable for this space?" | Noise compliance | → Noise Compliance |
| "Can people understand speech in this room?" | Speech intelligibility | → STI/SII Measurement |
| "How good does processed speech sound?" | Speech quality | → Speech Quality |
| "What's the SPL?" | Level measurement | → SPL Measurement |
| "What's the reverberation time?" | Room acoustics | → RT60 |
| "Is there a tonal component?" | Tonality | → Tonality |
2. Ask before computing
Before writing code, confirm these with the user:
Calibration: Do they have a CalibrationFactor? Did they calibrate with
calibrateMicrophone? (See Calibration section)Standard: Which standard or method applies? (ISO 532-1 vs 532-2, NC vs RC vs RNC)
Sound field: Free field or diffuse field recording?
Application context: What space type? What's the concern? (Drives NC vs RC vs RNC selection)
3. Select function and apply
Use the Key Functions table and the category-specific patterns below.
Functions That Do NOT Exist
| Function | Status | Use Instead |
|---|---|---|
pesq | Does NOT exist (proprietary ITU-T P.862 license) | visqol(degraded, reference, fs, Mode="speech") |
polqa | Does NOT exist (proprietary ITU-T P.863 license) | visqol(degraded, reference, fs) |
Never suggest pesq or polqa. They are not in Audio Toolbox and cannot be added due to licensing.
Key Functions
| Function | Purpose | Standard | Available From |
|---|---|---|---|
acousticLoudness | Perceived loudness (sones) | ISO 532-1/2 | R2020a |
acousticSharpness | Perceived sharpness (acum) | DIN 45692 | R2020a |
acousticRoughness | Perceived roughness (asper) | — | R2021a |
acousticFluctuation | Fluctuation strength (vacil) | — | R2020b |
integratedLoudness | Broadcast loudness (LUFS) | EBU R 128 / ITU-R BS.1770 | R2016b |
loudnessMeter | Streaming broadcast loudness meter | EBU R 128 / ITU-R BS.1770 | R2018a |
visqol | Full-reference audio quality (MOS) | — | R2024a |
stoi | Short-time objective intelligibility | — | R2024a |
mnru | Modulated noise reference unit | ITU-T P.810 | R2024a |
dBov | Power level relative to overload (dB) | ITU-T G.100.1 | R2024a |
speechTransmissionIndex | STI from IR or test signals | IEC 60268-16 | R2026a |
speechIntelligibilityIndex | SII from IR, test signals, or levels | ANSI/ASA S3.5-1997 | R2026b |
stipaExcitation | Generate STI/STIPA test signals | IEC 60268-16 | R2026a |
siiExcitation | Generate SII test signals | ANSI/ASA S3.5-1997 | R2026b |
noiseCriteria | NC rating + SIL + spectrum imbalance + rattle risk | ANSI/ASA S12.2 | R2026b |
roomNoiseCriteria | RNC rating (detects surging/fluctuations) | ANSI/ASA S12.2 | R2026b |
roomCriteria | RC Mark II rating (HVAC systems) | ANSI/ASA S12.2 | R2026b |
noiseCriteriaCurves | Look up NC curve values | ANSI/ASA S12.2 | R2026b |
noiseCriteriaRecommendations | Recommended NC levels by room type | ANSI/ASA S12.2 | R2026b |
speechInterferenceLevel | SIL from recording | ANSI/ASA S12.2 | R2026b |
rt60 | Reverberation time from impulse response | ISO 3382-1/2 | R2025a |
acousticToneToNoiseRatio | Detect/quantify tones | ECMA-418-1 | R2023b |
acousticProminenceRatio | Tonality via critical band comparison | ECMA-418-1 | R2023b |
splMeter | SPL measurement (time-weighted, per-band) | IEC 61672 | R2018a |
calibrateMicrophone | Compute calibration factor from known tone | — | R2018b |
weightingFilter | A/C/Z/K frequency weighting | IEC 61672-1 | R2016b |
octaveFilter | Single octave-band filter | ANSI S1.11-2004 | R2017b |
octaveFilterBank | Multi-band octave filtering | ANSI S1.11-2004 | R2019a |
See references/function-arguments.md for complete argument lists, defaults, and valid values for each function — consult when constructing calls with non-default parameters.
Calibration
CalibrationFactor from microphone sensitivity
Convert dB sensitivity (dBV/Pa) to the linear CalibrationFactor:
sensitivity_dBV = -26; % Example: -26 dBV/Pa microphone
calibrationFactor = 1 / 10^(sensitivity_dBV/20);The CalibrationFactor is the 3rd positional argument for acousticLoudness, acousticSharpness, acousticRoughness, and acousticFluctuation (default: sqrt(8)). For noise compliance functions, pass it as a name-value: CalibrationFactor=value.
System-level calibration advice
Always ask the user about their signal chain. If the recording path includes a preamp, audio interface, or ADC with unknown gain, the mic sensitivity spec alone is insufficient. Recommend:
Play a 1 kHz tone through the system and record it through the full signal chain (mic → preamp → ADC → WAV). Use a pistonphone for a known 94 dB reference, or place a physical SPL meter next to the microphone to get an SPL reading.
Compute the system calibration factor:
calibrationFactor = calibrateMicrophone(calRecording, fs, SPLreading, FrequencyWeighting="Z-weighting")Pass that factor to the measurement function
calibrateMicrophone expects a 1 kHz tone recording as input. The SPLreading argument is the known SPL in dB (e.g., 94 from a pistonphone, or the value shown on a physical SPL meter).
Cross-reference: for the physical recording workflow with calibrateMicrophone, see the matlab-play-record-audio skill.
Patterns
Psychoacoustic Metrics
[audioIn, fs] = audioread("recording.wav");
calFactor = 1 / 10^(-26/20);
loudness = acousticLoudness(audioIn, fs, calFactor, ...
Method="ISO 532-1", SoundField="diffuse", TimeVarying=true);
sharpness = acousticSharpness(audioIn, fs, calFactor, ...
SoundField="diffuse", TimeVarying=true);
roughness = acousticRoughness(audioIn, fs, calFactor, SoundField="diffuse");
fluctuation = acousticFluctuation(audioIn, fs, calFactor, SoundField="diffuse");Return types: acousticLoudness returns a scalar (sones) for stationary analysis. With TimeVarying=true, it returns a column vector (time-varying loudness in sones). The second output (specificLoudness) is a matrix (time x critical bands). It does NOT return a struct or timetable.
Ask the user:
ISO 532-1 (Zwicker, stationary/time-varying) or ISO 532-2 (Moore-Glasberg)?
Free-field or diffuse-field recording?
Time-varying analysis needed?
See references/psychoacoustic-metrics.md for method selection guidance (ISO 532-1 vs 532-2), interpretation tables (sone values, sharpness/roughness/fluctuation scales), tonality metric selection, and combined sound quality assessment workflows.
Unit Conversions
Use built-in functions — do NOT reimplement conversion formulas manually:
| Conversion | Function | Method selection |
|---|---|---|
| phon → sone | phon2sone(phon, standard) | "ISO 532-1" (default), "ISO 532-2" |
| sone → phon | sone2phon(sone, standard) | "ISO 532-1" (default), "ISO 532-2" |
| Hz → ERB | hz2erb(freq) | Single formula (no variants) |
| ERB → Hz | erb2hz(erb) | Single formula (no variants) |
| Hz → bark | hz2bark(freq) | Single formula (no variants) |
| bark → Hz | bark2hz(bark) | Single formula (no variants) |
| Hz → mel | hz2mel(freq, MelStyle=s) | "oshaughnessy" (default), "slaney" |
| mel → Hz | mel2hz(mel, MelStyle=s) | "oshaughnessy" (default), "slaney" |
When the user specifies a standard (e.g., "ISO 532-2"), pass it as the second positional argument for phon2sone/sone2phon. For hz2mel/mel2hz, use the MelStyle name-value pair.
Noise Compliance
Choose the function based on the application:
| Situation | Function | Why |
|---|---|---|
| General background noise rating | noiseCriteria | Standard NC curves (ANSI/ASA S12.2) |
| HVAC system noise specifically | roomCriteria | RC Mark II designed for HVAC evaluation |
| Noise with audible surging/fluctuation | roomNoiseCriteria | RNC penalizes time-varying noise (requires ≥15 s recording) |
| Quick speech interference check | speechInterferenceLevel | Direct SIL value |
| Need recommended levels for room type | noiseCriteriaRecommendations | Lookup table by room use |
[audioIn, fs] = audioread("office_noise.wav");
[NC, SIL, specImbalance, rattleRisk] = noiseCriteria(audioIn, fs, ...
CalibrationFactor=2.3);
[RC, LMF, QAI, rattleRisk] = roomCriteria(audioIn, fs, ...
CalibrationFactor=2.3);
SIL = speechInterferenceLevel(audioIn, fs, CalibrationFactor=2.3);
rec = noiseCriteriaRecommendations("NC", Occupancy="all");Return types:
NCis a string (e.g.,"NC-35"or"NC-56 (1000 Hz)"if a band exceeds)RCis a string (e.g.,"RC-32(N)"— the letter indicates spectrum quality: N/R/H/RV)SILandLMFare double scalars (dB)QAIis a double scalar (dB; 0 = ideal neutral)specImbalanceandrattleRiskare structs with.Summary(string) and.Details(table) — access.Summaryfor display. Exception:specImbalancedegrades to acharscalar (not a struct) when NC exceeds the curve range (e.g., "Above NC-70") or falls below NC-15. Guard withisstruct(specImbalance)before dot-indexing.noiseCriteriaRecommendationsrequires a criteria name argument:"NC","RNC", or"RC"
Ask the user:
What type of space? (office, classroom, hospital, concert hall)
Is the noise source HVAC? → prefer
roomCriteriaIs the noise steady or fluctuating? → if fluctuating, use
roomNoiseCriteria
See references/noise-compliance.md for detailed output interpretation and selection guidance.
Speech Quality
[degraded, fs] = audioread("processed_speech.wav");
[reference, ~] = audioread("reference_speech.wav");
mosScore = visqol(degraded, reference, fs, Mode="speech");
intelligibility = stoi(degraded, reference, fs);Critical: stoi and visqol argument order is (processed, reference, fs) — processed/degraded signal FIRST.
visqol modes: "speech" for narrowband/wideband speech (outputs 1-5 MOS), "audio" (default) for general audio.
STI/SII Measurement
Two methods are available for both speechTransmissionIndex and speechIntelligibilityIndex:
IR-based method (simpler — no alignment needed):
[ir, fs] = audioread("room_impulse_response.wav");
sti = speechTransmissionIndex(ir, fs);
sii = speechIntelligibilityIndex(ir, fs);Direct method (from test signals transmitted through the system):
fs = 48000;
reference = stipaExcitation(fs, Duration=15);
% Transmit reference through system, record as 'processed'
[processed, ~] = audioread("received_stipa.wav");
% IMPORTANT: Align signals before computing (see Signal Alignment)
processed = alignsignals(processed, reference);
sti = speechTransmissionIndex(processed, reference, fs);Ask the user:
Do they have an impulse response? → use IR method (simpler, no alignment needed)
Measuring live? → generate test signal, transmit, record, align, compute
STI or SII? STI for room/PA systems (IEC 60268-16); SII for hearing/communication (ANSI S3.5)
See references/sti-measurement.md for STI NV arguments, post-processing adjustment workflow (Annex M), test signal generation, and result interpretation. See references/sii-measurement.md for SII standard clause selection (5.1-5.3), BandImportanceFunction guidance, measurement positioning, and result interpretation.
SPL Measurement (Offline)
[audioIn, fs] = audioread("recording.wav");
spl = splMeter(SampleRate=fs, ...
Bandwidth="Full band", ...
FrequencyWeighting="A-weighting", ...
TimeWeighting="fast", ...
CalibrationFactor=2.3);
[Lt, Leq, Lpeak, Lmax] = spl(audioIn);
% For a single overall value, read the final row
fprintf("Leq: %.1f dB(A)\n", Leq(end));Valid Bandwidth values: "Full band" (overall level), "1 octave", "2/3 octave", "1/3 octave". There is no "broadband" option — use "Full band" for a single overall SPL value.
Output arity: All outputs (Lt, Leq, Lpeak, Lmax) have one row per input sample. For a 12-second file at 44.1 kHz, that is 539,648 rows. Never fprintf or disp the full output — it will hang. Instead:
Single overall value: read
Leq(end)(the final accumulated value)Time-interval readings: downsample to interval boundaries:
Leq(round(fs*T):round(fs*T):end, :)whereTis the desired interval in seconds
TimeInterval controls the integration window for Leq computation but does NOT reduce the number of output rows.
Do not manually compute 20*log10(rms(audioIn)/20e-6) — splMeter handles calibration, weighting, and time integration correctly per IEC 61672.
Multi-band (octave-band) SPL: Use splMeter with Bandwidth="1 octave" — do NOT use octaveFilterBank + manual dB conversion. Key properties for octave-band SPL:
[audioIn, fs] = audioread("recording.wav");
spl = splMeter(SampleRate=fs, ...
Bandwidth="1 octave", ...
FrequencyWeighting="Z-weighting", ...
TimeWeighting="fast", ...
FrequencyRange=[15 8000], ...
OctaveFilterOrder=8, ...
CalibrationFactor=2.3);
[Lt, Leq, Lpeak, Lmax] = spl(audioIn);
cf = getCenterFrequencies(spl);Critical gotchas for octave-band splMeter:
Property name is
OctaveFilterOrder(notFilterOrder).FilterOrderdoes not exist onsplMeter.FrequencyRangecompares against true center frequencies (fc = 1000 × 10^(3b/10)per IEC 61260), not nominal values. A band is included only if its true center falls within[lower, upper]. Bands below 1 kHz have true centers slightly above nominal (e.g., "63 Hz" = 63.10); bands above 1 kHz have true centers below nominal (e.g., "2 kHz" = 1995.26, "8 kHz" = 7943.28). Usefloor(true_center)for lower bounds andceil(true_center)for upper bounds. Common case "16 Hz to 8 kHz": use[15 8000]— the 16 Hz true center is 15.85 (so 15 ≤ 15.85 includes it) and 8 kHz true center is 7943 (so 8000 ≥ 7943 includes it). Do NOT use[16 8000]— this drops the 16 Hz band.Output is per-sample (see Output arity note above). Never print the full output. For time-interval readings, downsample:
Lt100ms = Lt(round(fs*0.1):round(fs*0.1):end, :);
See references/spl-measurement.md for weighting selection guidance (A/C/Z and fast/slow), SPL interpretation tables, occupational exposure limits, splMeter vs poctave equivalency, and common measurement scenarios.
Broadcast Loudness (EBU R 128)
[audioIn, fs] = audioread("broadcast_content.wav");
[loudness, loudnessRange] = integratedLoudness(audioIn, fs);
% Check compliance: target is -23 LUFS +/- 1 LU
isCompliant = abs(loudness - (-23)) <= 1;Use integratedLoudness for offline compliance checks. Use loudnessMeter for streaming/real-time monitoring (see matlab-process-streaming-audio).
RT60
[ir, fs] = audioread("room_ir.wav");
rtsummary = rt60(ir, fs, Bandwidth="1 octave");Ask the user:
What is the room type? (drives the target RT60)
Do they have an IR, or do they need to measure one?
T30 (standard) vs T20 (when noise floor limits dynamic range)?
See references/rt60-measurement.md for target RT60 values by space type, T20/T30/EDT interpretation, frequency-dependent analysis, and relationship to STI.
Tonality
[audioIn, fs] = audioread("machine_noise.wav");
[tnr, tnrFreq, isProminent] = acousticToneToNoiseRatio(audioIn, fs);
[pr, prFreq, isProminent] = acousticProminenceRatio(audioIn, fs);All outputs are empty (0×1) when no tonal components are detected. Check isempty(tnr) before indexing.
Standards-Compliant Filters (Custom Preprocessing)
The metric functions above perform standards-compliant filtering internally. Use standalone filter objects only for custom measurement workflows where no single metric function applies AND you do not need SPL output:
% A-weighting filter
wf = weightingFilter("A-weighting", fs);
weightedSignal = wf(audioIn);
% Octave-band filter bank (for waveform-level band decomposition only)
ofb = octaveFilterBank("1 octave", fs, ...
FrequencyRange=[31.5 16000], FilterOrder=8);
bandSignals = ofb(audioIn);If you need octave-band SPL levels, use splMeter with Bandwidth="1 octave" instead of octaveFilterBank + manual dB conversion. See the Multi-band SPL section above.
Signal Alignment
| Function | Internal alignment? | Action |
|---|---|---|
visqol | Yes | No action needed |
stoi | No | Align before calling |
speechTransmissionIndex | No | Align for direct method; IR method avoids issue |
speechIntelligibilityIndex | No | Align for direct method; IR method avoids issue |
For the direct (processed, reference, fs) method, always align signals:
processed = alignsignals(processed, reference);SII is more sensitive to misalignment than STI. At 50 ms delay: SII error ~0.016, STI negligible. At 200 ms: SII ~0.036, STI ~0.017. The IR-based (ir, fs) method avoids alignment issues entirely.
Common Mistakes
| Mistake | Why it fails | Correct approach |
|---|---|---|
Call pesq(reference, degraded, fs) | Function does not exist | visqol(degraded, reference, fs, Mode="speech") |
stoi(reference, processed, fs) | Wrong argument order | stoi(processed, reference, fs) — processed FIRST |
acousticFluctuationStrength(...) | Wrong function name | acousticFluctuation(audioIn, fs, calFactor) |
| Implement NC from scratch with hardcoded curves | Misses SIL, spectrum imbalance, rattle risk | noiseCriteria(audioIn, fs, CalibrationFactor=cf) |
20*log10(rms(x)/20e-6) for SPL | No weighting, no time integration, no calibration | splMeter with appropriate settings |
| Implement STI algorithm manually | ~100 lines, error-prone, misses NV options | speechTransmissionIndex(ir, fs) |
| Divide signal by calibration factor manually | Incorrect; factor is a multiplier, not a divisor | Pass as positional arg or CalibrationFactor=value |
| Skip alignment for direct STI/SII | Silent accuracy degradation | alignsignals(processed, reference) before calling |
Use acousticLoudness for broadcast loudness | Wrong standard (ISO 532 = sones, not LUFS) | integratedLoudness for EBU R 128 compliance |
string(rattleRisk) or fprintf("%s", specImbalance) | These are structs, not strings | Access .Summary field: rattleRisk.Summary |
specImbalance.Summary when NC is out of curve range | specImbalance degrades to char (not struct) when "Above NC-70" or below NC-15 | Guard: if isstruct(specImbalance), disp(specImbalance.Summary); end |
roomNoiseCriteria on recordings shorter than 15 s | ANSI/ASA S12.2-2019 requires ≥150 Lt samples at 100 ms intervals | Ensure recording is at least 15 seconds; use noiseCriteria for shorter clips |
round(NC) or arithmetic on NC/RC output | NC and RC are strings, not numbers | Use directly: fprintf("%s\n", NC) |
noiseCriteriaRecommendations() with no args | First argument (criteria name) is required | noiseCriteriaRecommendations("NC") |
loudness.OverallLoudness after acousticLoudness | Returns a scalar or vector, not a struct | Use directly: fprintf("%.1f sones\n", loudness) |
splMeter(..., FilterOrder=8) | Property does not exist on splMeter | Use OctaveFilterOrder=8 |
splMeter(..., FrequencyRange=[16 8000]) for 16 Hz–8 kHz | Drops 16 Hz band (true center is 15.85 Hz, which is below 16) | Use FrequencyRange=[15 8000] — use floor(true_center) for lower bound |
octaveFilterBank + 20*log10(rms(band)/pRef) for SPL | Manual, no time weighting, no calibration integration | Use splMeter with Bandwidth="1 octave" |
Treating splMeter output as a single value | All outputs have one row per input sample (e.g., 539k rows for 12s at 44.1 kHz) | Use Leq(end) for a single value, or downsample to interval boundaries |
Conventions
Always ask about the user's measurement goal before selecting a function
Always ask about calibration — recommend
calibrateMicrophonefor system-level accuracyDefault to
SoundField="free"for anechoic/outdoor,"diffuse"for reverberant roomsDefault to
Method="ISO 532-1"for loudness (Zwicker method, most common)Use IR-based methods for STI/SII when an impulse response is available (avoids alignment)
Pass
CalibrationFactorexplicitly rather than pre-scaling the signalFor noise compliance, prefer
roomCriteriafor HVAC andnoiseCriteriafor general background noise
Copyright 2026 The MathWorks, Inc.
Files
- ATTRIBUTION.mdAdded by AutoGineer1.1 KiB
- LICENSELicence file added by AutoGineer1.4 KiB
- SKILL.md22.7 KiB
- manifest.yaml444 B
- references/function-arguments.md7.3 KiB
- references/noise-compliance.md16.4 KiB
- references/psychoacoustic-metrics.md16.0 KiB
- references/rt60-measurement.md14.4 KiB
- references/sii-measurement.md16.9 KiB
- references/spl-measurement.md12.5 KiB
- references/sti-measurement.md14.0 KiB
references/sii-measurement.md
SII Measurement
Intake Questions
1. What is the application?
SII measures whether speech is audible and intelligible under given noise and hearing conditions. Common use cases:
| Application | Typical Approach |
|---|---|
| Room/system evaluation with a recording | IR-based or Direct method (like STI) |
| Hearing aid evaluation | Level-only method with InsertionGain and HearingThresholdLevel |
| Telecom or communication system design | Direct method through the system |
| Noise environment assessment (no audio) | Level-only method with estimated noise and speech levels |
| Classroom design for hearing-impaired students | Level-only method with HearingThresholdLevel and room noise |
2. Do they have audio, or just level specifications?
Have IR or audio: Use IR-based or Direct method (same as STI)
Only have noise levels and speech specs: Use the Level-only method -- no audio signal needed. Provide noise level, vocal effort, and distance.
3. Listener characteristics
Ask about:
Hearing loss: Does the listener have elevated hearing thresholds? ->
HearingThresholdLevel(dB HL per band)Conductive hearing loss: ->
ConductiveHearingLossLevel(must be <=HearingThresholdLevel)Hearing device: Is an amplification device being evaluated? ->
InsertionGain(gain curve per band)Binaural advantage: Is the listener using both ears? ->
IsBinaural=true(adds ~1 dB effective SNR benefit)
4. Communication conditions (Level-only method)
Vocal effort: Normal, raised, loud, or shout? (Changes spectral shape, not just level)
Distance: Talker-to-listener distance in meters (inverse-square-law attenuation applied)
Noise level: Measured or estimated background noise (scalar dB OASPL or per-band vector)
Measurement location: Free-field or at eardrum? (Affects the transfer function applied)
5. Band procedure
SII can use different frequency resolution:
"critical band"-- 21 bands (default for IR/Direct methods)"1/3 octave band"-- 18 bands"equally-contributing critical band"-- 17 bands"octave band"-- 6 bands (least precise, fastest)
The choice affects precision slightly; "1/3 octave band" is common for practical measurements.
Standard Clauses & Workflow Selection
The ANSI/ASA S3.5-1997 standard defines three calculation methods (clauses 5.1-5.3) in increasing order of generality. The standard also allows workflows outside these clauses.
When to Use Each Clause
| Clause | Conditions Required | Typical Scenario |
|---|---|---|
| 5.1 (Levels only) | Low reverberation; listener faces source directly (or omnidirectional); speech and noise are independent; system is linear | Office noise assessment, outdoor environments, simple room predictions |
| 5.2 (Measurements at listener position) | Well-mixed sound field OR direct facing/omnidirectional; linear system; identical conditions in both ears | Room acoustics with measurement mic at head center, PA system evaluation |
| 5.3 (Measurements at eardrum) | Linear system; identical conditions in both ears; measurements taken at eardrum (HATS or probe mic) | Hearing device evaluation, eardrum-level assessment, binaural research |
Practical Scenario Mapping
| Scenario | Recommended Approach | Rationale |
|---|---|---|
| "Will speech be intelligible in this room design?" (no measurements) | Clause 5.1 -- level-only method with estimated noise and standard speech | Fastest; no equipment needed; sufficient for design-stage predictions |
| "Evaluate speech intelligibility of installed PA system" | Clause 5.2 -- IR or direct method with mic at listener head position | Captures room acoustics and system response; standard positioning |
| "Assess hearing aid benefit for a patient" | Clause 5.3 -- with InsertionGain, HearingThresholdLevel, eardrum measurement location | Models device insertion gain and listener-specific hearing loss |
| "Compare noise environments for regulatory compliance" | Clause 5.1 -- with measured noise levels per band | Standard-compliant with measured noise; speech levels from standard |
| "Evaluate classroom acoustics with measured IR" | Clause 5.2 -- IR method with measured ambient noise | Realistic room assessment; compare to ANSI S12.60 targets |
| "Assess telecom system speech quality" | Clause 5.2 -- direct method with siiExcitation through the codec | Captures codec distortion and frequency shaping |
Measurement Positioning by Clause
Clause 5.1 (free-field, no audio):
No physical microphone measurement required for the SII computation itself
If measuring noise levels: place mic at center of listener's head position, without listener present
Speech levels typically come from the standard (selected via VocalEffort) or custom measurements at 1 m from talker
Clause 5.2 (free-field, at listener position):
Microphone at center of listener's head, mid-point between ears
Listener must NOT be present during measurement
Transmit excitation signal from the talker/source position
Set
SpeechMeasurementLocation="free-field"andNoiseMeasurementLocation="free-field"
Clause 5.3 (eardrum):
Use a Head and Torso Simulator (HATS) designed for eardrum measurements, OR
Probe tube microphone within 5 mm of eardrum on >=8 otologically normal subjects (age 18-30)
Set
SpeechMeasurementLocation="eardrum"andNoiseMeasurementLocation="eardrum"Must specify
SpeechAndNoiseLevelwhen using IR or direct method with eardrum location
Workflows Beyond the Standard
The three standard clauses do not cover every valid use case. Common non-standard workflows:
| Workflow | Approach | Key Differences from Standard Clauses |
|---|---|---|
| Hearing device evaluation (no room measurement) | Level-only method + InsertionGain + HearingThresholdLevel | Uses clause 5.1 inputs but adds device/listener parameters |
| Binaural advantage assessment | Any clause + IsBinaural=true | Reduces equivalent hearing threshold by 1.7 dB |
| Custom speech material weighting | Any clause + custom BandImportanceFunction vector | Weights bands differently than standard speech |
| Noise environment comparison (parametric) | Level-only method, vary NoiseLevel | Quick what-if analysis without re-measuring |
| In-ear monitor evaluation | Direct method + eardrum location + InsertionGain | Combines clause 5.3 positioning with device modeling |
| Classroom design for hearing-impaired | Level-only method + HearingThresholdLevel + room noise | Clause 5.1 with listener-specific hearing model |
BandImportanceFunction Selection
The BIF weights frequency bands by their relative importance to intelligibility. Choice depends on the speech material being evaluated:
| BIF Value | Speech Material | When to Use |
|---|---|---|
"standard" | Average English sentences | Default for most assessments -- general-purpose |
"NNS" | Nonsense syllables (CVC) | When testing pure audibility without linguistic context |
"CID-22" | CID Everyday Sentences | Clinical audiometry with sentence-level material |
"NU6" | Northwestern University Auditory Test No. 6 | Monosyllabic word recognition testing |
"DRT" | Diagnostic Rhyme Test | Military/tactical communication systems assessment |
"Short Passages" | Connected discourse (paragraphs) | Lecture/presentation intelligibility; high redundancy |
"SPIN" | Speech Perception in Noise sentences | Assessment in noisy environments; high-context and low-context |
| Custom vector | User-defined | Specialized material, non-English languages, music intelligibility |
Guidance:
Use
"standard"unless the assessment targets a specific test protocolFor hearing aid fitting/evaluation, match the BIF to the clinical test material used
For communication system certification, match the BIF to the test protocol (e.g., DRT for military standards)
Custom vectors must have one value per band (B values) and contain at least one nonzero element -- the function normalizes so they sum to 1
SpeechAndNoiseLevel for Standard Compliance
For IR-based and direct methods, the standard recommends specifying SpeechAndNoiseLevel. This corresponds to the Combined Speech and Noise Spectrum Level (CSNSL) in ANSI/ASA S3.5-1997 clauses 5.2/5.3 -- the unmodulated spectrum level of the received signal (speech + noise together) measured at the listener position or eardrum.
Measure with a calibrated SPL meter at the listener position during normal operational conditions (speech + noise simultaneously present)
Report as a per-band vector in dB SPL for standard compliance (scalar OASPL is accepted but may be non-compliant)
If you cannot measure combined levels, specify
NoiseLevelseparately (the function derives E' and N' differently)
When using the IR method with separately measured noise: The IR captures the room's effect on speech but does not contain ambient noise. If background noise is present during operation, specify NoiseLevel as a per-band vector measured at the listener position with the speech source off. The function adds this noise contribution to the SII calculation. Without it, the function assumes negligible noise (-50 dB/Hz per band).
speechIntelligibilityIndex
Three syntaxes are available. NV arguments are syntax-specific -- using the wrong argument for a syntax causes an error.
Method 1: IR-based
[ir, fs] = audioread("room_ir.wav");
sii = speechIntelligibilityIndex(ir, fs);Method 2: Direct (test signal)
fs = 48000;
[reference, speechRef] = siiExcitation(fs, Duration=16);
% --- Transmit 'reference' through system, record 'processed' ---
% Align signals (required -- SII is more sensitive to misalignment than STI)
% alignsignals requires vector input; use finddelay on one column instead
delay = finddelay(reference(:,1), processed(:,1));
if delay > 0
processed = processed(delay+1:end, :);
reference = reference(1:size(processed,1), :);
elseif delay < 0
reference = reference(-delay+1:end, :);
processed = processed(1:size(reference,1), :);
end
sii = speechIntelligibilityIndex(processed, reference, fs);Method 3: From levels only (no audio signal)
SII can be computed from speech and noise level specifications alone:
sii = speechIntelligibilityIndex();
sii = speechIntelligibilityIndex(Method="1/3 octave band", ...
NoiseLevel=40, VocalEffort="raised", Distance=2);Name-Value Arguments by Syntax
Valid for IR-based and Direct methods only:
| Argument | Default | Purpose |
|---|---|---|
SpeechAndNoiseLevel | -- | Combined speech+noise level per band (dB SPL vector). Recommended for standard compliance. |
NoiseLevel | -50 dB/Hz per band | Noise level -- numeric vector (dB SPL) or scalar (dB OASPL) |
Note: For standard-compliant SII results with the IR or Direct method, specify SpeechAndNoiseLevel as a per-band vector. Without it, the function uses default levels and emits a warning.
Valid for Level-only method only:
| Argument | Default | Purpose |
|---|---|---|
SpeechLevel | "standard" | "standard", "idealized", or numeric vector (dB SPL) |
NoiseLevel | -50 dB/Hz per band | Noise level -- numeric vector (dB SPL) or scalar (dB OASPL) |
Distance | 1 m | Talker-to-listener distance |
VocalEffort | "normal" | "normal", "raised", "loud", "shout" |
Valid for all syntaxes:
| Argument | Default | Purpose |
|---|---|---|
Method | -- | Band procedure: "critical band", "1/3 octave band", "equally-contributing critical band", "octave band" |
InsertionGain | 0 dB | Hearing device gain (scalar or 1-by-B vector) |
HearingThresholdLevel | 0 dB HL | Listener hearing threshold (scalar or 1-by-B vector) |
ConductiveHearingLossLevel | 0 dB HL | Conductive hearing loss (must be <= HearingThresholdLevel) |
BandImportanceFunction | "standard" | "standard", "NNS", "CID-22", "NU6", "DRT", "Short Passages", "SPIN" |
SpeechMeasurementLocation | "free-field" | "free-field" or "eardrum" |
NoiseMeasurementLocation | "free-field" | "free-field" or "eardrum" |
IsBinaural | false | Binaural listening advantage |
Critical constraints:
VocalEffort,Distance, andSpeechLevelare INVALID for IR-based and Direct methods -- they cause an errorSpeechAndNoiseLevelis INVALID for the Level-only method -- useSpeechLevelandNoiseLevelseparately insteadConductiveHearingLossLevelmust be <=HearingThresholdLevelin all bands
Ask the user about:
What is the talker-to-listener distance? (level-only method)
What vocal effort level? (level-only method)
Does the listener have hearing loss? (affects HearingThresholdLevel)
Is a hearing device being evaluated? (InsertionGain)
What noise level is present? (measured or estimated)
Outputs
[sii, baf, mtf] = speechIntelligibilityIndex(ir, fs, ...
NoiseLevel=40);sii-- Speech intelligibility index (0 to 1)baf-- Band audibility function (table)mtf-- Modulation transfer function (table)
siiExcitation
[mtfExcitation, speechExcitation, expLevels] = siiExcitation(fs, Duration=16);
mtfExcitation = siiExcitation(fs, Duration=20, ...
VocalEffort="raised", Method="1/3 octave band");The expLevels output provides expected speech spectrum levels for system calibration.
Signal Alignment
speechIntelligibilityIndex does not perform internal alignment for the direct method. SII is more sensitive to misalignment than STI.
Impact of misalignment (empirically measured):
| Delay | SII Error (20s signal) |
|---|---|
| <=50 ms | 0.016 (significant) |
| 100 ms | 0.029 |
| 200 ms | 0.036 |
Always align for the direct method. Because siiExcitation returns an N-by-9 matrix (not a vector), use finddelay on one column:
delay = finddelay(reference(:,1), processed(:,1));
if delay > 0
processed = processed(delay+1:end, :);
reference = reference(1:size(processed,1), :);
elseif delay < 0
reference = reference(-delay+1:end, :);
processed = processed(1:size(reference,1), :);
endThe IR-based method avoids this issue entirely -- impzest deconvolution handles propagation delay during IR extraction.
Complete Workflow: Evaluate SII for Hearing Device
sii = speechIntelligibilityIndex( ...
Method="1/3 octave band", ...
NoiseLevel=50, ...
VocalEffort="normal", ...
Distance=1, ...
InsertionGain=gainCurve, ...
HearingThresholdLevel=audiogram);Interpreting Results
SII Scale
| SII Range | Interpretation |
|---|---|
| 0.75-1.00 | Speech highly intelligible; nearly all speech cues audible |
| 0.45-0.75 | Moderate intelligibility; context helps significantly |
| 0.20-0.45 | Poor intelligibility; only familiar phrases understood |
| < 0.20 | Speech essentially inaudible above the noise |
Key Differences from STI
SII measures audibility of speech given noise and hearing -- can the listener detect speech cues?
STI measures transmission fidelity -- does the channel preserve modulations?
SII can be computed without any audio (level-only method) -- just noise spectrum + speech level + hearing threshold
SII accounts for listener hearing loss; STI does not
SII is more appropriate for hearing device evaluation; STI is more appropriate for room/PA evaluation
Band Audibility Function (BAF) Diagnostics
The BAF output shows the proportion of speech information audible in each frequency band (0 = completely masked, 1 = fully audible):
Low BAF at low frequencies: Low-frequency noise masks speech fundamentals -- less critical for intelligibility
Low BAF at 1-4 kHz: Critical concern -- these bands carry the most speech information (consonant cues)
Low BAF at high frequencies with normal hearing: HF noise or air absorption
Low BAF at high frequencies with hearing loss: Expected sensorineural pattern -- evaluate hearing device InsertionGain in these bands
Troubleshooting
| Observation | Likely Cause | Action |
|---|---|---|
| Error about invalid NV argument | Wrong argument for the syntax -- VocalEffort/Distance only valid for Level-only method | Check which syntax is being used; see constraints table above |
| SII = 1.0 (perfect) | Speech far above noise in all bands | Expected for close-range, quiet conditions with normal hearing |
| SII near 0 with moderate noise | HearingThresholdLevel too high, or noise level wrong | Verify units -- noise should be dB SPL, hearing in dB HL |
| SII doesn't change with InsertionGain | Gain doesn't overcome the hearing threshold deficit | Check that InsertionGain + SpeechLevel - NoiseLevel > HearingThresholdLevel in critical bands |
| Warning about default levels | SpeechAndNoiseLevel not specified for IR/Direct method | Provide measured per-band levels for standard-compliant results |
Copyright 2026 The MathWorks, Inc.
Requirements and compatibility
Works with: Claude Code, Claude apps, OpenAI Codex, GitHub Copilot, Cursor, Gemini CLI, Kiro, OpenCode, Goose, Amp, Roo Code, JetBrains Junie, Factory Droid, Devin (ex-Windsurf), Mistral Vibe, OpenHands, Letta
“Works with” means the tool reads the open Agent Skills format; a skill written for one tool may use features others lack.
Security
No issues found by our static scan· No issues found
A static scan is not a guarantee. Skills are third-party content: review a skill before you let an agent use it.
Attribution and licence
- Created by
- MathWorks
- Publisher
- MathWorks (Official)
- Source
- matlab/matlab-agentic-toolkit · skills-catalog/signal-processing/matlab-evaluate-acoustic-metrics · commit 2a7f215
- Licence
- BSD 2-Clause License (redistribution allowed)
- LICENSE (added by AutoGineer from LICENSE.md)
- ATTRIBUTION.md (written by AutoGineer)
- Synced
AutoGineer is not affiliated with or endorsed by MathWorks or MathWorks.
Download ZIP
The ZIP holds the skill’s files, its licence files and an ATTRIBUTION.md. Download matlab-evaluate-acoustic-metrics.zip